Correction of Fat-Water Swaps in Dixon MRI
Date
2016Author
Glocker, Ben
Konukoglu, Ender
Lavdas, Ioannis
Iglesias, Juan Eugenio
Aboagye, Eric O.
Rockall, Andrea G.
Rueckert, Daniel
Metadata
Show full item record
Glocker B. et al. (2016) Correction of Fat-Water Swaps in Dixon MRI. In: Ourselin S., Joskowicz L., Sabuncu M., Unal G., Wells W. (eds) Medical Image Computing and Computer-Assisted Intervention -- MICCAI 2016. MICCAI 2016. Lecture Notes in Computer Science, vol 9902. Springer, Cham
Abstract
The Dixon method is a popular and widely used technique for fat-water separation in magnetic resonance imaging, and today, nearly all scanner manufacturers are offering a Dixon-type pulse sequence that produces scans with four types of images: in-phase, out-of-phase, fat-only, and water-only. A natural ambiguity due to phase wrapping and local minima in the optimization problem cause a frequent artifact of fat-water inversion where fat- and water-only voxel values are swapped. This artifact affects up to 10 % of routinely acquired Dixon images, and thus, has severe impact on subsequent analysis. We propose a simple yet very effective method, Dixon-Fix, for correcting fat-water swaps. Our method is based on regressing fat- and water-only images from in- and out-of-phase images by learning the conditional distribution of image appearance. The predicted images define the unary potentials in a globally optimal maximum-a-posteriori estimation of the swap labeling with spatial consistency. We demonstrate the effectiveness of our approach on whole-body MRI with various types of fat-water swaps.