Shock decision algorithm for use during load distributing band cardiopulmonary resuscitation
View/ Open
Date
2021-08-15Author
Isasi Liñero, Iraia
Irusta Zarandona, Unai
Aramendi Ecenarro, Elisabete
Olsen, Jan-Age
Wik, Lars
Metadata
Show full item record
Resuscitation 165 : 93-100 (2021)
Abstract
Aim
Chest compressions delivered by a load distributing band (LDB) induce artefacts in the electrocardiogram. These artefacts alter shock decisions in defibrillators. The aim of this study was to demonstrate the first reliable shock decision algorithm during LDB compressions.
Methods
The study dataset comprised 5813 electrocardiogram segments from 896 cardiac arrest patients during LDB compressions. Electrocardiogram segments were annotated by consensus as shockable (1154, 303 patients) or nonshockable (4659, 841 patients). Segments during asystole were used to characterize the LDB artefact and to compare its characteristics to those of manual artefacts from other datasets. LDB artefacts were removed using adaptive filters. A machine learning algorithm was designed for the shock decision after filtering, and its performance was compared to that of a commercial defibrillator's algorithm.
Results
Median (90% confidence interval) compression frequencies were lower and more stable for the LDB than for the manual artefact, 80 min−1 (79.9–82.9) vs. 104.4 min−1 (48.5–114.0). The amplitude and waveform regularity (Pearson's correlation coefficient) were larger for the LDB artefact, with 5.5 mV (0.8–23.4) vs. 0.5 mV (0.1–2.2) (p < 0.001) and 0.99 (0.78–1.0) vs. 0.88 (0.55–0.98) (p < 0.001). The shock decision accuracy was significantly higher for the machine learning algorithm than for the defibrillator algorithm, with sensitivity/specificity pairs of 92.1/96.8% (machine learning) vs. 91.4/87.1% (defibrillator) (p < 0.001).
Conclusion
Compared to other cardiopulmonary resuscitation artefacts, removing the LDB artefact was challenging due to larger amplitudes and lower compression frequencies. The machine learning algorithm achieved clinically reliable shock decisions during LDB compressions.