dc.description.abstractWhen assessing the impact of novel concepts of scintillation block detectors for PET tomography by measuring the performance of single modules, or one module in coincidence with another, it is not trivial to translate the results of these investigations into a real PET scanner, made of several rings of modules. In particular, the real sensitivity, the signal to noise ratio of reconstructed images, and the spatial resolution (especially in the areas far from the center of field of view) among others are very critical to understand how good a PET scanner will perform for a real patient, and they are specific of the fully assembled scanner. Hence, demonstrating that a new idea under development is worth implementing in a real scanner requires more than just a measurement with two modules one in front of the other. For this reason, the set-up described in the present work will permit to reproduce full ring conditions using only a few modules at the price of a very good positioning accuracy.es_ES
dc.description.sponsorshipI must express all my gratitude towards one of my colleagues at CERN MME Group, Tommi Mikkola, for his infinite patience and teachings when reviewing the design and drawings of the mechanical parts.es_ES
dc.language.isoenges_ES
dc.publisherServicio Editorial de la Universidad del País Vasco/Euskal Herriko Unibertsitatearen Argitalpen Zerbitzuaes_ES
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.rights.urihttp://creativecommons.org/licenses/by-nc-sa/3.0/es/*
dc.titleDevelopment of a Test-Bench for the Accurate Positioning of Scintillation Detector Modules for Medical Imaging Applicationses_ES
dc.typeinfo:eu-repo/semantics/conferenceObjectes_ES
dc.rights.holderAtribución-NoComercial-CompartirIgual 3.0 España*
dc.departamentoesIngeniería de sistemas y automáticaes_ES
dc.departamentoesIngeniería mecánicaes_ES
dc.departamentoeuIngeniaritza mekanikoaes_ES
dc.departamentoeuSistemen ingeniaritza eta automatikaes_ES


Files in this item

Thumbnail
Thumbnail

This item appears in the following Collection(s)

Show simple item record

Atribución-NoComercial-CompartirIgual 3.0 España
Except where otherwise noted, this item's license is described as Atribución-NoComercial-CompartirIgual 3.0 España