Show simple item record

dc.contributor.authorBlanco, María Encarnación
dc.contributor.authorEncinas Pérez, Esther
dc.contributor.authorGonzález Mendia, Oscar ORCID
dc.contributor.authorRico Parrilla, Estitxu
dc.contributor.authorVozmediano, Valvanera
dc.contributor.authorSuárez González, María Elena
dc.contributor.authorAlonso Rojas, Rosa María ORCID
dc.date.accessioned2024-02-08T11:43:55Z
dc.date.available2024-02-08T11:43:55Z
dc.date.issued2015-03-09
dc.identifier.citationDrug Testing and Analysis 7(9) : 804-811 (2015)
dc.identifier.issn1942-7603
dc.identifier.urihttp://hdl.handle.net/10810/65715
dc.description.abstractIn this study, a selective and sensitive high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) method requiring low sample volume (≤100 μL) was developed and validated for the quantitative determination of the opioid drug fentanyl in plasma and cerebrospinal fluid (CSF). A protein precipitation extraction with acetonitrile was used for plasma samples whereas CSF samples were injected directly on the HPLC column. Fentanyl and 13C6-fentanyl (Internal Standard) were analyzed in an electrospray ionization source in positive mode, with multiple reaction monitoring (MRM) of the transitions m/z 337.0/188.0 and m/z 337.0/105.0 for quantification and confirmation of fentanyl, and m/z 343.0/188.0 for 13C6-fentanyl. The respective lowest limits of quantification for plasma and CSF were 0.2 and 0.25 ng/mL. Intra- and inter-assay precision and accuracy did not exceed 15%, in accordance with bioanalytical validation guidelines. The described analytical method was proven to be robust and was successfully applied to the determination of fentanyl in plasma and CSF samples from a pharmacokinetic and pharmacodynamic study in newborn piglets receiving intravenous fentanyl (5 µg/kg bolus immediately followed by a 90-min infusion of 3 µg/kg/h).es_ES
dc.language.isoenges_ES
dc.publisherWileyes_ES
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.titleQuantitative determination of fentanyl in newborn pig plasma and cerebrospinal fluid samples by HPLC-MS/MSes_ES
dc.typeinfo:eu-repo/semantics/articlees_ES
dc.rights.holderCopyright © 2015 John Wiley & Sons, Ltd*
dc.relation.publisherversionhttps://analyticalsciencejournals.onlinelibrary.wiley.com/doi/10.1002/dta.1778
dc.identifier.doi10.1002/dta.1778
dc.departamentoesQuímica analíticaes_ES
dc.departamentoeuKimika analitikoaes_ES
dc.identifier.eissn1942-7611